OWNER’S MANUAL
MicroTPI & MicroTPW
Field Portable Turbidimeters
HF scientific, inc.
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Fax: 239-332-7643
Toll free: 888-203-7248
Email: HFinfo@Watts.com
Website: www.hfscientific.com
Manual Part No. 24378 (5/10)
Rev. 1.9
Vista de página 0
1 2 3 4 5 6 ... 21 22

Resumo do Conteúdo

Página 1 - Field Portable Turbidimeters

OWNER’S MANUAL MicroTPI & MicroTPW Field Portable Turbidimeters HF scientific, inc. 3170 Metro Parkway Ft. Myers, FL 33916

Página 2

MicroTPI & MicroTPW (5/10) Rev. 1.9 41.3 The Touch Pad The touch pad has five buttons: ON/OFF, CAL, ↵, t, and u. The ON/OFF button is used

Página 3

MicroTPI & MicroTPW (5/10) Rev. 1.9 53.0 Normal Operation This instrument allows for the measurement of the turbidity of a grab sample. The

Página 4

MicroTPI & MicroTPW (5/10) Rev. 1.9 6 3.1 Routine Measurement Turn on the instrument by pressing the ON/OFF button for 1 second. The followi

Página 5 - Table of Contents

MicroTPI & MicroTPW (5/10) Rev. 1.9 74.0 Instrument Calibration The instrument was calibrated and tested prior to leaving the factory. Ther

Página 6

MicroTPI & MicroTPW (5/10) Rev. 1.9 84.3 Calibration Procedures The following procedure is recommended to perform a full range calibration.

Página 7 - Specifications

MicroTPI & MicroTPW (5/10) Rev. 1.9 98. Insert the 0.02 NTU standard into the sample well and press down until the cuvette snaps fully into

Página 8 - 1.0 Overview

MicroTPI & MicroTPW (5/10) Rev. 1.9 105.0 Routine Maintenance If the instrument is not kept in the supplied carrying case, when not in use,

Página 9

MicroTPI & MicroTPW (5/10) Rev. 1.9 11 5.2 Battery Replacement The instrument will periodically require new batteries to function properly.

Página 10

MicroTPI & MicroTPW (5/10) Rev. 1.9 126.0 Troubleshooting The instrument routinely performs self-diagnostics, and will automatically generat

Página 11 - 3.0 Normal Operation

MicroTPI & MicroTPW (5/10) Rev. 1.9 13 6.3 System Error Messages Normally, the cause of an error message is external to the instrument. If

Página 13

MicroTPI & MicroTPW (5/10) Rev. 1.9 14 7.0 Accessories and Replacement Parts List The items shown below are recommended accessories and rep

Página 14

MicroTPI & MicroTPW (5/10) Rev. 1.9 15 8.0 Warranty The manufacturer warrants to the original purchaser of this instrument that it will be f

Página 15 - ↵ button then release the

MicroTPI & MicroTPW (5/10) Rev. 1.9 169.0 Glossary Formazin Nephelometric Units (FNU): Equivalent in value to Nephelometric Turbidity Units

Página 16 - 5.0 Routine Maintenance

Micro TPI & MicroTPW (5/10) i Rev. 1.9 DECLARATION OF CONFORMITY Application of Council Directive Standard to which Conformity is Declared:

Página 18 - 6.0 Troubleshooting

Micro TPI & MicroTPW (5/10) ii Rev. 1.9 Table of Contents DECLARATION OF CONFORMITY...

Página 20

MicroTPI & MicroTPW (5/10) Rev. 1.9 1Specifications Measurement Range 0.01 –1100 NTU/FNU Accuracy ±2% of reading or 0.01 NTU (0-500 NTU) ±3

Página 21 - 8.0 Warranty

MicroTPI & MicroTPW (5/10) Rev. 1.9 2 1.0 Overview The MicroTPI/TPW allows for the measurement of turbidity in the field. The MicroTPI was

Página 22 - 9.0 Glossary

MicroTPI & MicroTPW (5/10) Rev. 1.9 3 Figure 1 is a depiction of the top of the instrument. The three main components of the instrument are