DRT-15CE (5/10)Page 5
7.0 CALIBRATION PROCEDURES
7.1 Calibration Standards
7.1.1 ProCal Standard Set (optional) Catalog No. 39071
HFProCalStandardsarerecommendedandcertiedbyHFscientic.
Theyaretraceabletofreshlypreparedformazinprimarystandards.These
standards are very easy to use off the shelf anytime without preparation
makingthemanidealturbiditystandard.ACerticateofTraceabilityis
availableonrequesttotheHFscienticCustomerServiceDepartment.HF
ProCal Standards may be used for calibration of HF turbidimeters. Order
fromHFscientic,inc.
NOTE: Donotfreezestandards.
Do not leave standards in the measuring well for extended
periods.
Do not shake standards.
Each ProCal Standard Kit contains:
-- 0.02 Calibration Standard
-- CertiedProCalStandards10.00,100.0,1000NTU
Standards are contained in preselected cuvettes with light
shield caps.
-- A sturdy storage case
7.1.2 Standard Formazin Solutions
CalibrationofthisinstrumentisbasedonFormazin,amaterialwhichis
madebypolymerization.
CalibrationsamplesmaybeobtainedbydilutingFormazinstocksuspension
using“Turbidity-Free”water.Formazinstocksuspensioncanbeprepared
by the user (Reference Standard Methods For Examination of Water and
Wastewater)orakitcanbepurchasedfrom,HFscientic,inc.,
Catalog No. 50040.